Z-1707-2017 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

Recall Details

Product Type
Devices
Report Date
April 12, 2017
Initiation Date
March 6, 2017
Termination Date
May 25, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2648

Product Description

Philips IntelliVue MX4O WLAN Patient Wearable Monitor Product: 865352 Exchange part (service numbers): 453564615311 TELE PWM,802.lla/b/g,ECG only, US only 453564615331 TELE PWM,802.1 1 a/b/g,ECG&Sp02, US only Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX4O is to be used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX4O is connected the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and alarming function. The MX4O does not automatically provide local monitoring or alarming when connected to the Information Center.

Reason for Recall

Philips lntelliVue MX4O WLAN Patient Wearable Monitor may not automatically switch to Monitor Mode with audible alarms when association with central monitoring system is unsuccessful [incomplete]

Distribution Pattern

Worldwide Distribution - US Nationwide including the countries of: Australia Belgium Canada Denmark Finland France Germany Germany India Italy Netherlands Norway Qatar Singapore South Africa Sweden Switzerland United Kingdom

Code Information

The following product number and Exchange part numbers containing the combination of HW Revision C.01 .01 with SW Revisions B.05 or B.06 of the Philips IntelliVue MX4O Patient Wearable Monitor