Z-1710-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- May 16, 2018
- Initiation Date
- October 19, 2017
- Termination Date
- April 24, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3
Product Description
BEQ-TOP 33500 3/8" ECC PACK, Catalog No. 701053752
Reason for Recall
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Distribution Pattern
Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
Code Information
Lot 3000056255