Z-1712-2017 Class III Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

Recall Details

Product Type
Devices
Report Date
April 12, 2017
Initiation Date
February 23, 2017
Termination Date
May 9, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
314 calibrators

Product Description

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c

Reason for Recall

Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.

Distribution Pattern

Worldwide Distribution - United States Nationwide: Denmark Sweden Norway Great Britain Belgium Netherlands Spain Israel Switzerland France Portugal Greece Germany Slovenia Czech Republic Hong Kong Colombia Korea Australia

Code Information

Lots 6JD079 and 6MD042