Z-1719-2014 Class II Terminated

Recalled by ConMed Corporation — Utica, NY

FDA device recall Z-1719-2014 was initiated by ConMed Corporation on May 8, 2014 and is designated Class II. Reason for recall: ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed. The recall status is terminated (terminated Apr 11, 2017). Affected quantity: Domestic - 12,600 units, Foreign - 44,280 units.

Recall details

Product Type
Devices
Report Date
Jun 11, 2014
Initiation Date
May 8, 2014
Termination Date
Apr 11, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Domestic - 12,600 units, Foreign - 44,280 units

Product description

MACROLYTE Premie Dispersive Electrode, REF 440-2400, CONMED CORPORATION. The device is applied to the patient during electrosurgical procedures to provide a path for RF current to leave the patient and return to the generator.

Reason for recall

ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed.

Distribution pattern

Worldwide Distribution - USA (nationwide) and the countries of Australia, Belgium, Brazil, Canada, Ecuador, El Salvador, Finland, France, Great Britain, Honduras, India, Indonesia, Israel, Italy, Japan, Latvia, Lebanon, Malaysia, Mexico, Panama, Philippines, Portugal, Saudi Arabia, South Africa, Spain, Thailand, Trinidad, United Arab Emirates and Uruguay.

Code information

Beginning Lot Code 1204275; Ending Lot Code 1401075 (Lot code indicates the date of manufacture = yymmdds, where s is the shift).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1719-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for ConMed Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1719-2014. Report a data issue: hello@regdatalab.com.