Z-1732-2013 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-1732-2013 was initiated by Terumo Cardiovascular Systems Corporation on Feb 27, 2013 and is designated Class II. Reason for recall: Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for… The recall status is terminated (terminated Feb 17, 2016). Affected quantity: 3972 since 4/26/12.

Recall details

Product Type
Devices
Report Date
Jul 24, 2013
Initiation Date
Feb 27, 2013
Termination Date
Feb 17, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3972 since 4/26/12

Product description

CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.

Reason for recall

Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations.

Distribution pattern

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV and the countries of Jordan, CANADA, AUSTRALIA, Japan ,Hong Kong ,UNITED ARAB EMIRATES (UAE), BELGIUM, South Korea, Malaysia, Mexico, Philippines, Singapore, Taiwan, and Thailand.

Code information

Catalog number: 500AV* and serial numbers: 0024, 0033, 0040, 0051, 1192, 1193, 1355,1356, 1420, 1423, 1464, 1465, 1467, 1468,1483,1504,1521-1524,1526,1527,1529,1530, 1602-1604, 1606, 1662, 2023, 2035, 2102-2107, 2800, 2901,2902, 3131-3135, 3261, 3262, 4132, 5330-5333, 5338-5340, 5578, and 5579.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1732-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-1732-2013. Report a data issue: hello@regdatalab.com.