Z-1737-2013 Class II Terminated

Recalled by SpineFrontier, Inc. — Beverly, MA

FDA device recall Z-1737-2013 was initiated by SpineFrontier, Inc. on Dec 30, 2011 and is designated Class II. Reason for recall: One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles. The recall status is terminated (terminated Jan 8, 2014). Affected quantity: 2 rods.

Recall details

Product Type
Devices
Report Date
Jul 24, 2013
Initiation Date
Dec 30, 2011
Termination Date
Jan 8, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 rods

Product description

KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients.

Reason for recall

One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.

Distribution pattern

USA Nationwide Distribution including the state of TX

Code information

Lot Number: 002876-007R

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1737-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1737-2013. Report a data issue: hello@regdatalab.com.