Z-1738-2015 Class II Terminated

Recalled by Smith & Nephew, Inc. — Memphis, TN

FDA device recall Z-1738-2015 was initiated by Smith & Nephew, Inc. on May 6, 2015 and is designated Class II. Reason for recall: The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell. The recall status is terminated (terminated Apr 1, 2022). Affected quantity: 40 units.

Recall details

Product Type
Devices
Report Date
Jun 17, 2015
Initiation Date
May 6, 2015
Termination Date
Apr 1, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 units

Product description

R3(TM) 28 MM I.D., 60 MM O.D. CONSTRAINED LINER, REF 71339160, QTY: (1), STERILE EO, Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.

Reason for recall

The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain

Code information

Lot numbers: 16HM08087, 14HM08097, 14HM08133, 14HM08141, and 14JM21517

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1738-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Smith & Nephew, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Smith & Nephew, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Smith & Nephew, Inc. →

Source: openFDA record for Z-1738-2015. Report a data issue: hello@regdatalab.com.