Z-1746-2017 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

FDA device recall Z-1746-2017 was initiated by Becton Dickinson & Company on Mar 16, 2017 and is designated Class II. Reason for recall: BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit… The recall status is terminated (terminated May 1, 2018). Affected quantity: 13,064,500 tubes (non-expired).

Recall details

Product Type
Devices
Report Date
Apr 12, 2017
Initiation Date
Mar 16, 2017
Termination Date
May 1, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,064,500 tubes (non-expired)

Product description

BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations.

Reason for recall

BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the user removes the needle from the stopper following specimen collection.

Distribution pattern

Worldwide - US Nationwide Distribution

Code information

Catalog Number: 367716 - Lot numbers: 5251673 with expiry 3/31/2017; 6064542 with expiry 9/30/17; 6187520 with expiry 1/31/2018 and 6253669 with expiy 3/31/18 Catalog Number: 369714 - Lot numbers: 5272777 with expiry 4/30/17; 5336909 with expiry 6/30/17; 5357632 with expiry 7/31/17; 6040953 and 6064634 both with expiry 9/30/17; 6090572 with expiry 10/31/17; 6124932 with expiry 11/30/17; 6173983 with expiry 12/31/17; 6187600 with expiry 1/31/18 and 6216656 with expiry 2/28/18 Lot numbers 6064542, 6187520 and 6253669 - limited to distribution outside the U.S. market. UDI for lot 6253669 is DI (01)30382903677161, PI (17)180331(10)6253669(30)0100 and lot 6216656 is DI (01)30382903697145, PI (17)180228(10)6216656(30)0100; for all other lots UDI is not applicable based on manufacture date.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1746-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1746-2017. Report a data issue: hello@regdatalab.com.