Z-1753-2013 Class II Terminated

Recalled by Leibel-Flarsheim Company LLC — Cincinnati, OH

Recall Details

Product Type
Devices
Report Date
July 24, 2013
Initiation Date
April 25, 2013
Termination Date
July 26, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product Description

Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.

Reason for Recall

A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.

Distribution Pattern

Nationwide Distribution (US), including the states of TN, KY, CO, KS, CT, HI, and AZ.

Code Information

The following catalog numbers, serial numbers and associated manufacturing dates are affected by this recall: 700539 CI0807H273 August2007; 700539 CI0106H162 January 2006; 700539 CI1010H523 October 2010; 700539 CI0811H567 August 2011; 700539 CI0108H313 January 2008; 700539 CI0309H418 March 2009; 700539 15031229 Apri12003; 700539 14031227 Apri12003; 700539 CI0404H045 January 2005; 700539 CI0504H054 May 2004; 700539 CI1106H208 November 2006; 700539 CI0509H424 May 2009