Z-1753-2024 Class II
Recall Details
- Product Type
- Devices
- Report Date
- May 8, 2024
- Initiation Date
- March 7, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 41 devices
Product Description
INNOMED CUP REMOVAL STARTER INSTRUMENT, REF 520054, Hip cutter blade cup
Reason for Recall
54 MM finisher blade was inadvertently affixed to the shaft of the CupX handle in lieu of the 54 MM starter blade.
Distribution Pattern
FL, GA, PA, NY, CA, VA, SD, MN, TX, UT, NC, and UK, Denmark, Germany Switzerland, Ireland, Canada
Code Information
UDI/DI 00840277104007, Lot Number 1223K