Z-1753-2025 Class II Ongoing

Recalled by Richard Wolf GmbH — Knittlingen

Recall Details

Product Type
Devices
Report Date
May 21, 2025
Initiation Date
April 9, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
31 units

Product Description

PiezoWave 2 Control Unit, Product Number 100506US, compact myofascial acoustic compression therapy device that enables treatment with piezoelectric shockwaves (Therapeutic massager).

Reason for Recall

The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.

Distribution Pattern

US State: GA

Code Information

UDI-DI: 00814334022863; Serial Numbers: 5000553232 5000553233 5000553234 5000553235 5000553236 5000553237 5000553238 5000553239 5000553240 5000553241 5000553242 5000553243 5000553244 5000553245 5000553246 5000553247 5000553248 5000553249 5000553250 5000553251 5000553482 5000553483 5000553484 5000553485 5000553486 5000553488 5000553502 5000553503 5000553504 5000553505 5000553506