Z-1758-2024 Class II Ongoing

Recalled by SPINEART SA — Plan-Les-Ouates, N/A

FDA device recall Z-1758-2024 was initiated by SPINEART SA on Mar 15, 2024 and is designated Class II. Reason for recall: Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is… The recall status is ongoing. Affected quantity: 242 systems.

Recall details

Product Type
Devices
Report Date
May 15, 2024
Initiation Date
Mar 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
242 systems

Product description

PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).

Reason for recall

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Distribution pattern

U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.

Code information

Model Numbers/Lot-Batch Numbers: TLF-DS 45 25-S 7-4407 TLF-DS 45 25-S 7-6207 TLF-DS 45 30-S 7-4244 TLF-DS 45 30-S 7-6226 TLF-DS 45 35-S 7-4242 TLF-DS 45 35-S 7-5470 TLF-DS 45 35-S 7-6684 TLF-DS 45 40-S 7-6232 TLF-DS 45 45-S 7-5474 TLF-DS 45 45-S 7-6874 TLF-DS 55 30-S 7-4409 TLF-DS 55 30-S 7-4695 TLF-DS 55 30-S 7-4965 TLF-DS 55 35-S 7-4413 TLF-DS 55 35-S 7-6455 TLF-DS 55 40-S 7-4415 TLF-DS 55 40-S 7-4417 TLF-DS 55 45-S 7-4411 TLF-DS 55 45-S 7-5055 TLF-DS 55 45-S 7-6087 TLF-DS 55 45-S 7-6459 TLF-DS 65 35-S 7-4405 TLF-DS 65 35-S 7-6230 TLF-DS 65 35-S 7-6461 TLF-DS 65 35-S 7-6882 TLF-DS 65 40-S 7-4248 TLF-DS 65 40-S 7-4250 TLF-DS 65 40-S 7-5480 TLF-DS 65 40-S 7-6228 TLF-DS 65 40-S 7-6554 TLF-DS 65 40-S 7-6740 TLF-DS 65 45-S 7-4582 TLF-DS 65 45-S 7-5484 TLF-DS 65 45-S 7-5808 TLF-DS 65 45-S 7-6668 TLF-DS 65 45-S 7-6742 TLF-DS 65 45-S 7-7299 TLF-DS 65 50-S 7-5488 TLF-DS 65 50-S 7-9264 UDI-DI for associated Model Numbers: TLF-DS 45 25-S 07640185345799 TLF-DS 45 30-S 07640185345805 TLF-DS 45 35-S 07640185345812 TLF-DS 45 40-S 07640185345829 TLF-DS 45 45-S 07640185345836 TLF-DS 55 30-S 07640185345850 TLF-DS 55 35-S 07640185345867 TLF-DS 55 40-S 07640185345874 TLF-DS 55 45-S 07640185345881 TLF-DS 60 35-S 07640305166082

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1758-2024
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for SPINEART SA? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1758-2024. Report a data issue: hello@regdatalab.com.