Z-1768-2019 Class I Terminated

Recalled by CareFusion 303, Inc. — San Diego, CA

FDA device recall Z-1768-2019 was initiated by CareFusion 303, Inc. on Apr 13, 2018 and is designated Class I. Reason for recall: The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as… The recall status is terminated (terminated Dec 26, 2023). Affected quantity: 607618.

Recall Details

Product Type
Devices
Report Date
Jul 24, 2019
Initiation Date
Apr 13, 2018
Termination Date
Dec 26, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
607618

Product Description

Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following products are affected by this recall as they are pumps or pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic. Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203), and Alaris Pump module Bezel Kit Assembly (P/N 10964559 and P/N 49000204). Product Usage: The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural. There are 6 bezel posts on the Alaris Pump bezel assembly.

Reason for Recall

The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.

Distribution Pattern

Worldwide Distribution - U.S. Nationwide CA, TX, NC, AL, ID, TN, IN, MA, UT, HI, NH, PA, IL, OR, FL, MI, VA, AR, OH, GA, NY, MD, MN, NM, AZ, KY, WI, NE, MS, DC, MO, KS, SC, LA, MT, WA, NJ, CO, CT, OK, AK, AS, CU, DE, GU, IA, ME, MP, ND, NV, PR, SD, VT, WV, WY; OUS: AE, AU, BH, CA, CH, CN, GI, IL, IN, KW, MX, MY, NL, NZ, PH, PK, QA, SA, SG, TW, and ZA.

Code Information

All pumps manufactured between April 2011 and June 2017, that have bezel mechanical assemblies manufactured with FR-110 plastic resin. And all pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1768-2019
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.