Z-1772-2017 Class II Terminated
Recalled by Draegar Medical Systems, Inc.
— Andover, MA
FDA device recall Z-1772-2017 was initiated by Draegar Medical Systems, Inc. on Mar 28, 2017 and is designated Class II. Reason for recall: It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit. The recall status is terminated (terminated Sep 17, 2018). Affected quantity: 2,156 (total for all devices).
Recall details
- Product Type
- Devices
- Report Date
- Apr 19, 2017
- Initiation Date
- Mar 28, 2017
- Termination Date
- Sep 17, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,156 (total for all devices)
Product description
Delta, Catalog Number: MS18597 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Product Usage: The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patients medical condition. The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be used when the patient is not under gas anesthesia. For combination with Scio gas module: Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the In-finity monitors.
Reason for recall
It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
Distribution pattern
Worldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of: AR AT BE BG BN BO BY CA CH CL CO CZ DE EG ES FI FR GB GR HR HU ID IE IL IN IT LT MX MY NL NO NP PA PE PK PL PT RO RS RU SA SE SK TH TR TW UG UZ VN ZA
Code information
VF10.0 software
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1772-2017
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Draegar Medical Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works