Z-1777-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 19, 2019
- Initiation Date
- May 20, 2019
- Termination Date
- December 15, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4 units
Product Description
Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86662003 Product Usage: Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Reason for Recall
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Distribution Pattern
Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine
Code Information
Serail Numbers: WAA088731, WAA089124 , WAA089128, WAA089586