Z-1777-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 29, 2020
- Initiation Date
- October 9, 2019
- Termination Date
- December 11, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 53,790 units total
Product Description
Medivators EndoGator REF 100651. Kit for Olympus 140/160/180/190 Series GI Endoscopes for CO2 Connection with Endogator EGP-100, Olympus OFP, Olympus AFU-100 & Erbe EIP 2 Irrigation Pump Units. Contents 10. Sterile EO. of 100145CO2EXT Endo SmartCap Irrigation Tubing for Olympus 140/160/180/190 Series GI Endoscopes with Extended CO2 Input - Product Usage: is intended to be used with an air or CO2 source and/or a pump along with a sterile water source to supply air or CO2 and sterile water to an endoscope during endoscopic procedures.
Reason for Recall
Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.
Distribution Pattern
Worldwide distribution - US Nationwide distribution and the countries of Australia, England.
Code Information
Lot # / Case UDI#s: 430358 / 2067796406530322060110430358, 429884 / 2067796406530322060110429884, 428806 / 2067796406530322070110428806, 430568 / 2067796406530322060110430568, 431254 / 2067796406530322070110431254, 431455 / 2067796406530322070110431455, 431603 / 2067796406530322070110431603, 431834 / 2067796406530322070110431834, 431985 / 2067796406530322070110431985, 431973 / 2067796406530322070110431973, 432922 / 2067796406530322070110432922, 432108 / 2067796406530322070110432108, 432921 / 2067796406530322070110432921, 433116 / 2067796406530322070110433116, 433115 / 2067796406530322080110433115.