Z-1793-2012 Class II Terminated
FDA device recall Z-1793-2012 was initiated by Aesculap Implant Systems on Feb 21, 2012 and is designated Class II. Reason for recall: During an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of the extension stem. The recall status is terminated (terminated Jan 15, 2014). Affected quantity: 8.
Recall details
- Product Type
- Devices
- Report Date
- Jun 20, 2012
- Initiation Date
- Feb 21, 2012
- Termination Date
- Jan 15, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8
Product description
Reason for recall
Distribution pattern
Nationwide Distribution including PA and TX.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report