Z-1802-2026 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 22, 2026
- Initiation Date
- March 19, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,848 units
Product Description
B. Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 2.6MM PUMP SEGMENT. Model Number: B3-3632M3705.
Reason for Recall
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Canada.
Code Information
Model Number: B3-3632M3705; UDI-DI Primary: 04046955348824; UDI-DI Unit: 04046955348817; All lots manufactured since 01AUG2025.