Z-1804-2020 Class II

Recalled by Boston Scientific Corporation — Marlborough, MA

FDA device recall Z-1804-2020 was initiated by Boston Scientific Corporation on Sep 18, 2019 and is designated Class II. Reason for recall: Certain iLab Polaris Systems which were upgraded with a previous 2.10 software version (disk batch 629774) contained a bug in which only the True Color Secondary Capture (SC) format will archive… Affected quantity: 150 units.

Recall details

Product Type
Devices
Report Date
May 6, 2020
Initiation Date
Sep 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
150 units

Product description

iLab Polaris Multi-Modality Guidance System; iLab Ultrasound Imaging System

Reason for recall

Certain iLab Polaris Systems which were upgraded with a previous 2.10 software version (disk batch 629774) contained a bug in which only the True Color Secondary Capture (SC) format will archive Diastolic Hyperemia-Free Ratio data as intended. If either Multi-Frame UltraSound (US) or SC Image Storage formats are used, the DFR data will not be archived and the user will not be notified of this unexpected behavior.

Distribution pattern

OUS: Japan, Belgium, Denmark, France, Great Britain, and Spain. No U.S. consignees.

Code information

Polaris 2.10 software version (disk batch 629774); iLab Polaris Systems affected product listing (serial numbers): 200204, 200209, 200217, 200308, 200310, 200448, 200450, 200456, 200457, 200458, 200461,200480, 200481, 200103, 200145, 200044, 200118, 200360, 200498, 200170, 8361, 10312, 10443, 10548, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278,10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278, 10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 7402, 8397, 8497, 8498, 8508, 8516, 8522, 8523, 8530, 10044, 10053, 10275, 10278, 10362, 10365, 10496, 10500, 10565, 10566, 10604, 10651, 10667, 10694, 10753, 10770, 10867, 10872, 10873, 10875, 10876, 10952, 8475, 6582 , 7023, 7857, 5637, 5716, 6650, 6894, 4031, 5003, 5043, 5060, 5109, 5232, 5239, 5393, 5465, 5651, 5678, 5860, 6193, 6350, 8264, 6819, 6921, 6934, 7332, 7341, 7362, 7733, 7828, 7858, 7933, 7934, 8011, 8155, 8290, 6383, 5182, 5219, 5296, 5398, 8157, 6760, 7723, 10108, 8365, 8367, 10504, 10657, 10501, 4274, 8046, 8048, 5892, 7818, 4773, 5585, 8113, 5165, 4402, 6468, 6319, 5313, 7511, 4952, 5277, 5457, 4116, 5403, 5078, 5664, 5666, 6327, 6733, 6994, 6136, 6206, 6172, 7604, 7602, 7353, 6273, 6993, 4045, 10221, 10261, and 8540.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1804-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Boston Scientific Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Boston Scientific Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Boston Scientific Corporation →

Source: openFDA record for Z-1804-2020. Report a data issue: hello@regdatalab.com.