Z-1805-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
May 23, 2018
Initiation Date
August 7, 2017
Termination Date
August 23, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
842 products in total

Product Description

Persona Stemmed Tibial Provisional Size H Left, Item Number 42532108301 For use during orthopaedic surgery for implantation of a prosthesis.

Reason for Recall

There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.

Distribution Pattern

Worldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NJ, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI, and Puerto Rico; and to countries of: Australia, Canada, Hong Kong, India, Japan, Korea, Malaysia, Netherlands, and Singapore.

Code Information

Lot 63535352, UDI (01)00889024246935(10)63607357; Lot 63535352, UDI (01)00889024246935(10)63535352