Z-1810-2025 Class I Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

Recall Details

Product Type
Devices
Report Date
June 11, 2025
Initiation Date
April 22, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1853 total

Product Description

Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C

Reason for Recall

Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.

Distribution Pattern

US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.

Code Information

UDI-DI 10888277858060 (EA); 40888277858061 (CASE) KIT LOTS 24CBJ925 24DBT405 24FBO318