Z-1811-2025 Class I Ongoing

Recalled by MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL

FDA device recall Z-1811-2025 was initiated by MEDLINE INDUSTRIES, LP - Northfield on Apr 22, 2025 and is designated Class I. Reason for recall: Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate. The recall status is ongoing. Affected quantity: 1,853 total.

Recall details

Product Type
Devices
Report Date
Jun 11, 2025
Initiation Date
Apr 22, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,853 total

Product description

Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU DYNJ59270 2. CRANI PACK-LF SKU DYNJ44805M 3. CRANIOTOMY SKU CDS982719V CDS982719W DYNJ904168F DYNJ908404B DYNJ908723A DYNJ908723C 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A 5. CRANIOTOMY CDS SKU CDS981888X CDS983467I 6. CRANIOTOMY PACK-LF SKU DYNJ09882G DYNJ09882I 7. CRANIOTOMY S NASSAU SKU DYNJ905879G 8. MHC CRANIOTOMY SKU DYNJ903713K DYNJ903713L 9. PAD CRANIOTOMY SKU DYNJ902149J

Reason for recall

Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.

Distribution pattern

US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.

Code information

1. CRANI ACCESSORY PACK SKU DYNJ59270 UDI-DI 10888277771512 (ea), 40888277771513 (case) Lots 23JDB683 23JDC548 24ADB783 24HDA822 2. CRANI PACK-LF SKU DYNJ44805M UDI-DI 10195327506872 (ea), 40195327506873 (case) Lots 23JBI747 23KBL228 24ABG696 24CBM117 24EBQ114 24GBJ699 3. CRANIOTOMY SKU CDS982719V UDI-DI 10195327510831 (ea), 40195327510832 (case) Lots 23KBP461 23LBN381 24ABB119 24BBC547 24BBI145 SKU CDS982719W UDI-DI 10195327645090 (ea), 40195327645091 (case) Lots 24DBS813 SKU DYNJ904168F UDI-DI 10195327395728 (ea), 40195327395729 (case) Lots 24DBU188 SKU DYNJ908404B UDI-DI 10195327410506 (ea), 40195327410507 (case) Lots 23IDB680 23LDB546 24BDB174 24FDA741 24FDB584 24IDA225 24IDA646 24IDA840 SKU DYNJ908723A UDI-DI 10195327403270 (ea), 40195327403271 (case) Lots 23IBO314 23JBC218 24ABK090 SKU DYNJ908723C UDI-DI 10198459009785 (ea), 40198459009786 (case) Lots 24IBM205 24IBN365 25ABM356 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A UDI-DI 10195327466466 (ea), 40195327466467 (case) Lots 23JLA765 24DLA724 24DLB006 24ELA325 5. CRANIOTOMY CDS SKU CDS981888X UDI-DI 10195327144760 (ea), 40195327144761 (case) Lots 24ABT432 24CBD698 24HBH954 24HBP483 24IBG797 25CBF098 SKU CDS983467I UDI-DI 10195327457945 (ea), 40195327457946 (case) Lots 23KDA336 6. CRANIOTOMY PACK-LF SKU DYNJ09882G UDI-DI 10195327574055 (ea), 40195327574056 (case) Lots 24CBF778 SKU DYNJ09882I UDI-DI 10198459058172 (ea), 40198459058173 (case) Lots 24HBN426 7. CRANIOTOMY S NASSAU SKU DYNJ905879G UDI-DI 10195327512996 (ea), 40195327512997 (case) Lots 23KBI086 23LBH647 23LBH649 24BBU253 24DBV133 24FBE409 8. MHC CRANIOTOMY SKU DYNJ903713K UDI-DI 10195327201579 (ea), 40195327201570 (case) Lots 23JBK426 SKU DYNJ903713L UDI-DI 10195327531218 (ea), 40195327531219 (case) Lots 23KBT939 23LBS862 24ABH577 24CBH126 24CBH771 9. PAD CRANIOTOMY SKU DYNJ902149J UDI-DI 10195327554323 (ea), 40195327554324 (case) Lots 23KBP275 24ABJ684 24CBF216 24DBI296 24EBD959 24FBF064 24HBK164 24IBN443

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1811-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for MEDLINE INDUSTRIES, LP - Northfield? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for MEDLINE INDUSTRIES, LP - Northfield: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about MEDLINE INDUSTRIES, LP - Northfield →

Source: openFDA record for Z-1811-2025. Report a data issue: hello@regdatalab.com.