Z-1819-2013 Class II Terminated

Recalled by Custom Spine, Inc. — Parsippany, NJ

FDA device recall Z-1819-2013 was initiated by Custom Spine, Inc. on Mar 28, 2013 and is designated Class II. Reason for recall: On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required… The recall status is terminated (terminated Oct 24, 2013). Affected quantity: 18 units.

Recall details

Product Type
Devices
Report Date
Aug 7, 2013
Initiation Date
Mar 28, 2013
Termination Date
Oct 24, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18 units

Product description

Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus Dr. Parsippany, NJ 07054 USA. TEL: 973-808-0019 The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine for pedicular and non-pedicular fixation.

Reason for recall

On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm.

Distribution pattern

Nationwide Distribution including Nebraska and Virginia.

Code information

Catalog Number 048545; Lot Number L363

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1819-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Custom Spine, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1819-2013. Report a data issue: hello@regdatalab.com.