Z-1820-2017 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-1820-2017 was initiated by Philips Electronics North America Corporation on Mar 22, 2017 and is designated Class II. Reason for recall: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in… The recall status is terminated (terminated May 17, 2021). Affected quantity: 6992 (total for all devices).

Recall details

Product Type
Devices
Report Date
Apr 26, 2017
Initiation Date
Mar 22, 2017
Termination Date
May 17, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6992 (total for all devices)

Product description

Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039

Reason for recall

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Distribution pattern

Nationwide, worldwide including Canada

Code information

Devices with Laird Chiller.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1820-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1820-2017. Report a data issue: hello@regdatalab.com.