Z-1820-2017 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

Recall Details

Product Type
Devices
Report Date
April 26, 2017
Initiation Date
March 22, 2017
Termination Date
May 17, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6992 (total for all devices)

Product Description

Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039

Reason for Recall

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Distribution Pattern

Nationwide, worldwide including Canada

Code Information

Devices with Laird Chiller.