Z-1821-2017 Class II Terminated
FDA device recall Z-1821-2017 was initiated by Philips Electronics North America Corporation on Mar 22, 2017 and is designated Class II. Reason for recall: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in… The recall status is terminated (terminated May 17, 2021). Affected quantity: 6992 (total for all devices).
Recall details
- Product Type
- Devices
- Report Date
- Apr 26, 2017
- Initiation Date
- Mar 22, 2017
- Termination Date
- May 17, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6992 (total for all devices)
Product description
UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039
Reason for recall
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Distribution pattern
Nationwide, worldwide including Canada
Code information
Devices with a Laird Chiller.