Z-1821-2017 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-1821-2017 was initiated by Philips Electronics North America Corporation on Mar 22, 2017 and is designated Class II. Reason for recall: Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in… The recall status is terminated (terminated May 17, 2021). Affected quantity: 6992 (total for all devices).

Recall details

Product Type
Devices
Report Date
Apr 26, 2017
Initiation Date
Mar 22, 2017
Termination Date
May 17, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6992 (total for all devices)

Product description

UNIQ: UNIQ FD10 722026; UNIQ FD10/10 722027; UNIQ FD20/10 biplane 722029; UNIQ FD10 OR Table 722033; UNIQ FD10/10 OR Table 722034; UNIQ FD20/10 OR Table 722036; UNIQ FD20/20 722038; UNIQ FD20/20 biplane OR Table 722039

Reason for recall

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Distribution pattern

Nationwide, worldwide including Canada

Code information

Devices with a Laird Chiller.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1821-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1821-2017. Report a data issue: hello@regdatalab.com.