Z-1822-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 26, 2017
- Initiation Date
- March 22, 2017
- Termination Date
- May 17, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6992 (total distributed for all products)
Product Description
The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031
Reason for Recall
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Distribution Pattern
Nationwide, worldwide including Canada
Code Information
Devices with a Laird Chiller.