Z-1823-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 7, 2013
- Initiation Date
- June 10, 2013
- Termination Date
- December 19, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 200
Product Description
Cardinal Steam Biological Indicator, SteriTec Product Code PL 218, Distributor Product Code T40360EZ EZTest Steam Biological Indicator: Mesa EZTest Steam Biological Indicator is a self contained Biological Indicator intended for monitoring the efficacy of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. Mesa EZTest has a recommended incubation time of 24 hours.
Reason for Recall
Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.
Distribution Pattern
Worldwide distribution: US (Nationwide) and countries of: Argentina, Australia, Canada, British Virgin Island, Czech Republic, Italy, and Turkey.
Code Information
Affected Biological Indicator Lot No.: S-444