Z-1836-2014 Class II Terminated

Recalled by Claymount Assembli — Dinxperlo

Recall Details

Product Type
Devices
Report Date
July 9, 2014
Initiation Date
March 15, 2014
Termination Date
October 18, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 devices

Product Description

The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system.

Reason for Recall

Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases.

Distribution Pattern

Worldwide Distribution - USA including NC, OH, CO, CA, NJ, IL, OH, PA, FL and Internationally to Canada, China, France, Italy, Netherlands, South Africa, Switzerland, Turkey, and United Kingdom.

Code Information

*** US Devices Only *** 1) Part ID: 17360; Serial Numbers: WO186722/E-1 , WO196470/E-1 , WO198534/E-1. 2) Part ID: 17361; Serial Numbers: WO186723/E-1. 3) Part ID: 17362; Serial Numbers: WO186724/E-1 , WO196360/E-1 , WO198545/E-2. 4) Part ID: 18476; Serial Numbers: WO196188/E-1 , WO196188/E-2 , WO196189/E-2. 5) Part ID: 18477; Serial Numbers: WO198547/E-1. 6) Part ID: 18479; Serial Numbers: WO198552/E-1.