Z-1841-2014 Class II Terminated

Recalled by SeaSpine Inc — Vista, CA

Recall Details

Product Type
Devices
Report Date
July 2, 2014
Initiation Date
May 21, 2014
Termination Date
December 15, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7

Product Description

Integra UCR 3mm Hex, Large Axial Driver. The UCR Hex Axial Driver is a surgical instrument used to implant and remove polyaxial screws of varying reduction heights provided with the Malibu", UCR, and NewPort" Systems. The Malibu", UCR, NewPort" Systems are intended to be used as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur. Accepted standard techniques of spinal fusion are an integral part of the implantation of these systems. Indications For Use: The Malibu", UCR, NewPort" Systems are intended to be used as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur. Accepted standard techniques of spinal fusion are an integral part of the implantation of these systems. The intended use of the Malibu, UCR, and NewPort Systems, when used as a Pedicle Screw Spinal System or Spondylolisthesis Spinal Fixation Device System, is to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine.

Reason for Recall

Integra has identified through an investigation of complaints that there may be the potential for the tip of a single lot number of the UCR 3mm Hex, Large Axial Driver, 91-1123 to twist or become rounded during use.

Distribution Pattern

US Distribution including the states of : AL, CA, OH, NM and TN.

Code Information

Catalogue Number: 91- 1123. Lot No. M016090B.