Z-1841-2016 Class II Terminated

Recalled by Integra York PA, Inc — York, PA

FDA device recall Z-1841-2016 was initiated by Integra York PA, Inc on Mar 11, 2016 and is designated Class II. Reason for recall: One lot of self-lock chisel blade handles have a wider slot that holds the chisel blade that may result in the inability to firmly secure the blade in the handle. The recall status is terminated (terminated Apr 17, 2017). Affected quantity: 678 Chisel Blade Holders.

Recall details

Product Type
Devices
Report Date
Jun 8, 2016
Initiation Date
Mar 11, 2016
Termination Date
Apr 17, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
678 Chisel Blade Holders

Product description

Miltex Premium Grade Self-Lock Chisel Blade Handle, Catalogue No. 4-405, manual surgical instrument.

Reason for recall

One lot of self-lock chisel blade handles have a wider slot that holds the chisel blade that may result in the inability to firmly secure the blade in the handle.

Distribution pattern

Worldwide distribution. US nationwide, Mexico, and Korea.

Code information

Catalogue No. 4-405, Lot # 97327

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1841-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Integra York PA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Integra York PA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Integra York PA, Inc →

Source: openFDA record for Z-1841-2016. Report a data issue: hello@regdatalab.com.