Z-1843-2014 Class II Terminated

Recalled by BioFire Diagnostics, Inc. — Salt Lake City, UT

FDA device recall Z-1843-2014 was initiated by BioFire Diagnostics, Inc. on May 27, 2014 and is designated Class II. Reason for recall: BioFire has identified an increased risk of false positive results when the FilmArray Blood Culture Identification (BCID) Panel is used with bioM¿rieux BacT/ALERT Standard Anaerobic (SN) blood culture… The recall status is terminated (terminated Nov 7, 2014). Affected quantity: 1084.

Recall details

Product Type
Devices
Report Date
Jul 2, 2014
Initiation Date
May 27, 2014
Termination Date
Nov 7, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1084

Product description

FirmArray Blood Culture Identification (BCID) Panel, Model 2.0 The FilmArray Blood Culture Identification (BCID) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test designed to test positive blood cultures to quickly identify the organism that is growing in the blood culture. The following gram-positive bacteria, gram-negative bacteria, and yeast are identified using the FilmArray BCID Panel: Enterococci, Listeria monocytogenes, commonly encountered Staphylococci (including specific differentiation of Staphylococcus aureus), commonly encountered Streptococci (with specific differentiation of Streptococcus agalactiae, Streptococcus pneumoniae, and Streptococcus pyogenes), Acinetobacter baumannii, commonly encountered Enterobacteriaceae (including specific differentiation of the Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and Serratia marcescens), Haemophilus influenzae, Neisseria meningitidis (encapsulated), Pseudomonas aeruginosa, Candida albicans, Candida glabrata, Candida krusei, Candida parapsilosis, and Candida tropicalis. The FilmArray BCID Panel also contains assays for the detection of genetic determinants of resistance to methicillin (mecA), vancomycin (vanA and vanB), and carbapenems (blaKPC) to aid in the identification of potentially antimicrobialresistant organisms in positive blood culture samples.

Reason for recall

BioFire has identified an increased risk of false positive results when the FilmArray Blood Culture Identification (BCID) Panel is used with bioM¿rieux BacT/ALERT Standard Anaerobic (SN) blood culture bottles. False positive results have been observed for Pseudomonas aeruginosa and Enterococcus.

Distribution pattern

Worldwide Distribution - USA (nationwide) European Union, Hong Kong, and VA.

Code information

426614,424514,424014,421114,419114,419214,417814,416214,416114,411214,411314,411014,410414,410014,408414,406314,404314,402314,403613,700113,698513,695313,692813,689813,689113,686313,685313,677713,668613,665013,667213,667113,663313,663013,665913,657313,659613,651413,647513,644613,632713,424614,424114,419514,419614,418014,416314,416414,412014,412114,411914,409914,406214,404414,700013,698413,695413,692913,689213,685413,668713,665113,667313,662913,657213,663913,660213,651513,647313,622213,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1843-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for BioFire Diagnostics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1843-2014. Report a data issue: hello@regdatalab.com.