Z-1861-2012 Class II Terminated

Recalled by Stryker Medical Division Of Stryker Corporation — Portage, MI

FDA device recall Z-1861-2012 was initiated by Stryker Medical Division Of Stryker Corporation on May 17, 2012 and is designated Class II. Reason for recall: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells… The recall status is terminated (terminated May 12, 2014). Affected quantity: 999.

Recall details

Product Type
Devices
Report Date
Jul 4, 2012
Initiation Date
May 17, 2012
Termination Date
May 12, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
999

Product description

Stryker Rose Bed Model (FL14E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone

Reason for recall

An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac

Distribution pattern

Worldwide distribution:USA (nationwide) and countries including: Argentina, Bertec, Brasil, Brazil, Canada, Chile, China, India, Latin America, Mexico,Osteonics SA, Polska, Portugal, Puerto Rico, and Singapore.

Code information

Model FL14E: Serial Number QB22371, QB22370, QB22369, QB22368, QB22367, QB21200, QB18330, QB17733, B22524, B22523, B22522, B22521, B22506, B22505, B22504, B22503, B22502, B22501, B22500, B22446, B22445, B22444, B22443, B22442, B22441, B22440, B22439, B22438, B22437, B22436, B22435, B22427, B22426, B22425, B22424, B22423, B22422, B22421, B22420, B22419, B22373, B22372, B22308, B22307, B22306, B22305, B22304, B22303, B22302, B22301, B22300, B22299, B22298, B22297, B22296, B22295, B22294, B22293, B22292, B22291, B22290, B22289, B22247, B22246, B22245, B22244, B22243, B22242, B22241, B22240, B22239, B22238, B22237, B22236, B22235, B22234, B22233, B22232, B22231, B22230, B22229, B22228, B22227, B22226, B22225, B22224, B22223, B22222, B22221, B22220, B22219, B22218, B22217, B22216, B22215, B22214, B22213, B22212, B22211, B22210, B22209, B22208, B22207, B22206, B22205, B22204, B22203, B22202, B22201, B22200, B22199, B22198, B22197, B22196, B22195, B22194, B22193, B22192, B22191, B22190, B22189, B22188, B22187, B22186, B22185, B22184, B22172, B22171, B22170, B22169, B22168, B22167, B22166, B22165, B22164, B22163, B22162, B22161, B22160, B22159, B22158, B22157, B22156, B22120, B22119, B22118, B22117, B22116, B22115, B22114, B22113, B22112, B22111, B22110, B22109, B22108, B22107, B22106, B22105, B22104, B22103, B22102, B22101, B22100, B22099, B22098, B22097, B22096, B22095, B22094, B22093, B22092, B22091, B22090, B22089, B22088, B22087, B22086, B22085, B22084, B22083, B22082, B22081, B22080, B22079, B22078, B22077, B22076, B22075, B22074, B22073, B22072, B22071, B22070, B22069, B22068, B22067, B22066, B22065, B22064, B22063, B22062, B22061, B22060, B22059, B22058, B22057, B22056, B22055, B22054, B22053, B22052, B22051, B22050, B22049, B22048, B22047, B22046, B22045, B22044, B22043, B22042, B22041, B22040, B22039, B22038, B22037, B22036, B22035, B22034, B22033, B22032, B22031, B22030, B22029, B22028, B22027, B22026, B22025, B22024, B22023, B22022, B22021, B22020, B22019, B22018, B22017, B22016, B22015, B22014, B22013, B22012, B21975, B21974, B21973, B21972, B21971, B21970, B21969, B21968, B21967, B21966, B21965, B21964, B21963, B21962, B21961, B21960, B21959, B21958, B21898, B21897, B21896, B21895, B21894, B21893, B21892, B21891, B21890, B21810, B21809, B21808, B21807, B21806, B21805, B21804, B21803, B21787, B21786, B21785, B21784, B21783, B21782, B21781, B21780, B21779, B21744, B21743, B21742, B21741, B21740, B21739, B21738, B21737, B21736, B21735, B21734, B21733, B21732, B21731, B21730, B21729, B21728, B21727, B21726, B21725, B21724, B21723, B21722, B21721, B21720, B21719, B21718, B21717, B21716, B21715, B21714, B21713, B21712, B21711, B21710, B21709, B21708, B21707, B21706, B21705, B21704, B21703, B21702, B21701, B21700, B21699, B21698, B21697, B21696, B21695, B21694, B21634, B21633, B21632, B21631, B21630, B21629, B21628, B21627, B21626, B21625, B21624, B21623, B21622, B21621, B21620, B21619, B21618, B21595, B21594, B21593, B21592, B21591, B21590, B21589, B21588, B21587, B21586, B21585, B21584, B21583, B21582, B21581, B21580, B21579, B21578, B21577, B21576, B21575, B21574, B21549, B21514, B21513, B21512, B21501, B21500, B21499, B21498, B21497, B21496, B21495, B21494, B21493, B21473, B21472, B21471, B21470, B21469, B21468, B21467, B21466, B21465, B21464, B21463, B21385, B21384, B21383, B21382, B21381, B21380, B21379, B21378, B21348, B21347, B21346, B21345, B21344, B21343, B21342, B21341, B21340, B21339, B21338, B21337, B21336, B21335, B21334, B21316, B21315, B21314, B21313, B21312, B21302, B21301, B21300, B21299, B21298, B21297, B21296, B21295, B21294, B21293, B21292, B21291, B21290, B21207, B21206, B21205, B21204, B21203, B21202, B21201, B21199, B21198, B21197, B21196, B21195, B21189, B21188, B21187, B21186, B21185, B21184, B21183, B21182, B21181, B21180, B21143, B21142, B21141, B21140, B21139, B21138, B21137, B21136, B21135, B21134, B21133, B21132, B21131, B21075, B21074, B21073, B21072, B21071, B21070, B20969, B20968, B20967, B20966, B20965, B20964, B20963, B20962, B20961, B20960, B20959, B20958, B20957, B20956, B20955, B20954, B20943, B20942, B20941, B20940, B20939, B20938, B20937, B20936, B20935, B20934, B20933, B20932, B20931, B20930, B20929, B20928, B20927, B20926, B20925, B20924, B20923, B20922, B20921, B20920, B20919, B20918, B20917, B20916, B20915, B20914, B20913, B20912, B20911, B20910, B20909, B20908, B20907, B20906, B20905, B20904, B20903, B20902, B20901, B20900, B20899, B20898, B20897, B20896, B20895, B20894, B20893, B20892, B20891, B20890, B20889, B20866, B20865, B20764, B20763, B20762, B20761, B20760, B20759, B20758, B20757, B20756, B20755, B20754, B20753, B20752, B20751, B20750, B20749, B20748, B20747, B20746, B20745, B20744, B20676, B20675, B20674, B20673, B20672, B20671, B20670, B20669, B20589, B20588, B20587, B20586, B20585, B20584, B20583, B20582, B20581, B20580, B20579, B20578, B20577, B20576, B20575, B20513, B20512, B20511, B20465, B20444, B20443, B20442, B20441, B20440, B20439, B20438, B20437, B20436, B20435, B20434, B20266, B20265, B20264, B20263, B20184, B20183, B20182, B20181, B20180, B20179, B20178, B20177, B20176, B20175, B20174, B20173, B20172, B20171, B20170, B20169, B20168, B20167, B20166, B20165, B20002, B20001, B20000, B19999, B19998, B19997, B19996, B19995, B19994, B19993, B19992, B19991, B19591, B19590, B19589, B19588, B19587, B19586, B19585, B19584, B19583, B19431, B19430, B19429, B19428, B19427, B19426, B19425, B19424, B19302, B19301, B19300, B19299, B19298, B19297, B19296, B19295, B19294, B19293, B19292, B19291, B19290, B19289, B19288, B19287, B19286, B19285, B19284, B19283, B19282, B19281, B19280, B19279, B19278, B19277, B19276, B19275, B19274, B19273, B19272, B19271, B19270, B19269, B19268, B19267, B19266, B19265, B19264, B19263, B19262, B19261, B19260, B19259, B19258, B19257, B19256, B19255, B19254, B19253, B19252, B19251, B19250, B19249, B19248, B19247, B19246, B19245, B19244, B19243, B19242, B19241, B19240, B19239, B19238, B19237, B19236, B19235, B19234, B19233, B19232, B19231, B19230, B19125, B19124, B19123, B19122, B19121, B19120, B19119, B19118, B19117, B19116, B19115, B19114, B19113, B19112, B19111, B19110, B19109, B19108, B19107, B19106, B19105, B19104, B19103, B19102, B19101, B19100, B19099, B19098, B19097, B19096, B19066, B19065, B19064, B19063, B19062, B19061, B19060, B19059, B19058, B19057, B19056, B19055, B19054, B19053, B19052, B19051, B19050, B19049, B19048, B19047, B19046, B19045, B19044, B19043, B19042, B19041, B19040, B19039, B19038, B19037, B19036, B18984, B18983, B18982, B18981, B18980, B18978, B18977, B18976, B18975, B18974, B18952, B18951, B18949, B18948, B18947, B18941, B18884, B18883, B18882, B18881, B18880, B18879, B18878, B18877, B18876, B18875, B18874, B18873, B18872, B18835, B18834, B18833, B18832, B18831, B18830, B18829, B18828, B18827, B18826, B18825, B18824, B18823, B18822, B18821, B18820, B18819, B18706, B18705, B18704, B18703, B18702, B18701, B18700, B18699, B18698, B18697, B18696, B18695, B18694, B18693, B18692, B18691, B18690, B18689, B18688, B18687, B18686, B18685, B18684, B18683, B18670, B18669, B18668, B18667, B18666, B18575, B18574, B18573, B18561, B18560, B18488, B18487, B18486, B18485, B18484, B18483, B18482, B18481, B18480, B18479, B18478, B18477, B18476, B18475, B18474, B18331, B18205, B18204, B18203, B18202, B17830, B17829, B17824, B17759, B17739, B17738, B17737, B17736, B17735, B17734, B17683, B17682, B17681, B17680, B17679, B17678, B17677, B17676, B17675, B17616, B17615, B17614, B17613, B17612, B17611, B17610, B17461, B17334, B17333, B17332, B17331, B17330, B17295, B17294, B17293, B17292, B17179, B17178, B17177, B16999, B16998, B16997, B16996, B16995, B16994, B16934, B16933, B16932, B16931, B16731, B16730, B16729, B16728, B16727, B16726, B16725, B16724, B16723, B16702, B16701, B16700, B16699, B16698, B16697, B16696, B16695, B16694, B16693, B16692, B16691, B16690, B16689, B16682, B16641, B16640, B16639, B16638, B16637, B16341, B16340, B16339, B16338, B16015, B16014, B16013, B16012, B16011, and B16010

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1861-2012
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1861-2012. Report a data issue: hello@regdatalab.com.