Z-1871-2013 Class II Terminated

Recalled by SpineFrontier, Inc. — Beverly, MA

FDA device recall Z-1871-2013 was initiated by SpineFrontier, Inc. on Nov 9, 2012 and is designated Class II. Reason for recall: Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical… The recall status is terminated (terminated Jun 5, 2015). Affected quantity: 13 drill guides.

Recall details

Product Type
Devices
Report Date
Aug 14, 2013
Initiation Date
Nov 9, 2012
Termination Date
Jun 5, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 drill guides

Product description

Drill Guide - Chameleon FacetFuse Fixation System, Model: 11-00028 Rev. A FacetFuse Drill Guides were designed to provide a concentric pathway for the drills and taps to access the bone while protecting the soft tissue from the cutting geometry of the drills/taps. In addition the drill guide is used as a reference to measure depth that the drills and taps are protruding into bone, which is measured by depth markings at the proximal end of the drills/taps relative to the proximal end of the drill guide. These devices were used during implantation of spinal intervertebral body fusion devices intended for intervertebral body fusion of the spine of skeletally mature patients.

Reason for recall

Prior to use in surgery, the Drill Guides were presenting noticeable degrees of denting and/or bending. As a result, drills and taps could not be directed down guides during surgery due to mechanical interference.

Distribution pattern

US Distribution including Puerto Rico and the states of CO, KS, MA and TX.

Code information

2520301, 2520307

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1871-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1871-2013. Report a data issue: hello@regdatalab.com.