Z-1871-2016 Class II Terminated

Recalled by Skytron, Div. The KMW Group, Inc — Grand Rapids, MI

FDA device recall Z-1871-2016 was initiated by Skytron, Div. The KMW Group, Inc on Mar 28, 2016 and is designated Class II. Reason for recall: Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury. The recall status is terminated (terminated Jul 25, 2017). Affected quantity: 24.

Recall details

Product Type
Devices
Report Date
Jun 8, 2016
Initiation Date
Mar 28, 2016
Termination Date
Jul 25, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
24

Product description

Ergon III Series Skyboom Monitor Bracket. Part # 3FCM1 Flatscreen Bracket, single monitor mount, 27inch max. screen size

Reason for recall

Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury.

Distribution pattern

Worldwide distribution. US states: VA, VT, UT, GA, NJ, and IL; Japan.

Code information

Model Number(s): 3FCM1; Manufactured Date: September 2015; Serial Numbers 1507T719-00-0-0074, 1507T719-00-0-0091, 1507T719-00-0-0095, 1507T719-00-0-0096, 1507T719-00-0-0102, 1507T719-00-0-0107, 1508T719-00-0-0129, 1507T719-00-0-0087, 1507T719-00-0-0069, 1507T719-00-0-0103, 1507T719-00-0-0104, 1507T719-00-0-0066, 1507T719-00-0-0072, 1507T719-00-0-0073, 1507T719-00-0-0075, 1507T719-00-0-0088, 1507T719-00-0-0101, 1507T719-00-0-0108, 1507T719-00-0-0112, 1507T719-00-0-0113, 1507T719-00-0-0090, 1507T719-00-0-0097, 1507T719-00-0-0099, and 1507T719-00-0-0100

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1871-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Skytron, Div. The KMW Group, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Skytron, Div. The KMW Group, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Skytron, Div. The KMW Group, Inc →

Source: openFDA record for Z-1871-2016. Report a data issue: hello@regdatalab.com.