Z-1875-2012 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-1875-2012 was initiated by Terumo Cardiovascular Systems Corporation on May 9, 2012 and is designated Class II. Reason for recall: During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips… The recall status is terminated (terminated Apr 24, 2013). Affected quantity: 53,139 (each) in total.

Recall details

Product Type
Devices
Report Date
Jul 4, 2012
Initiation Date
May 9, 2012
Termination Date
Apr 24, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
53,139 (each) in total

Product description

Sarns Antegrade Cardioplegia Cannula: ROOT INFUSTION VENT CATHETER WITH 12 GAUGE BLUE NEEDLE RECTANGULAR FLANGE, 6" (15 CM) LONG Product Usage: The Sarns Cardioplegia Cannula and Cardioplegia/Vent Catheters are indicated for placement in the aortic root to deliver cardioplegia solution or blood to the heart, to vent the left heart, and to aspirate air from the aorta during cardiopulmonary bypass procedures.

Reason for recall

During production of the Sarns Antegrade Cardioplegia Cannula, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of a foreign substance on the inner surface of some cannulae tips. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface. The substance was likely deposited during the molding process, but the exact composition of the

Distribution pattern

Worldwide Distribution - US (nationwide) including the states of: AL, AR , AZ , CA, CT , FL, IA , IL, IN , KS , LA , MI , NM, MO , NC, NE , NJ , NV, NY , OH, OK , PA, RI , SC , TN, TX, WA , WI, WV and the countries of: BELGIUM, BRAZIL, Byelorussian SSR, CAMBODIA , Dominican Republic, Honduras, Japan, Malaysia, Mexico, Nicaragua, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, UNITED ARAB EMIRATES (UAE), and Vietnam.

Code information

Catalog number: 203861: 0548731, 0552411, 0556018, 0558941, 0563005, 0565138, 0565841, 0569745, 0570058, 0571163, 0571584, 0573043, 0573288, 0575518, 0580832, 0583446,0585251, 0586160, 0590448, 0591942, 0596210, 0600490, 0607020, 0618429, 0619201, 0630155, 0634777, 0640509, 0649185, 0652212, 0655898, 0661142, and 0668560.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1875-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-1875-2012. Report a data issue: hello@regdatalab.com.