Z-1880-2013 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-1880-2013 was initiated by Stryker Instruments Div. Of Stryker Corporation on Jun 24, 2013 and is designated Class II. Reason for recall: A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the… The recall status is terminated (terminated Jan 30, 2014). Affected quantity: 215 units total.

Recall details

Product Type
Devices
Report Date
Aug 14, 2013
Initiation Date
Jun 24, 2013
Termination Date
Jan 30, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
215 units total

Product description

Sonopet Ultrasonic Aspirator Console 5450-852-000 (Console 230V) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.

Reason for recall

A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali

Distribution pattern

Worldwide - USA Nationwide including the states of CA, CO, FL, HI, IA, IL, KS, KY, LA, MD, MI, MN, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, ITALY, and the countries of AUSTRALIA, CANADA, UNITED KINGDOM, NETHERLANDS, INDIA, JAPAN, NEW ZEALAND, SWITZERLAND, HONG KONG, and TAIWAN.

Code information

5450-852-000 (Console 230V) with SN: 1134000114, 1206800484 1210900274, 1134000124, 1206800494, 1210900284,1134000134 1206800504, 1210900294, 1134000144, 1206800514, 1210900304, 1134000154, 1206800524, 1211800064, 1134000164, 1206800534, 1211800074, 1134000174, 1206800544, 1211800084, 1134000184, 1208700014, 1211800094, 1134000194, 1208700024, 1212900014, 1134000204, 1208700034, 1213100114, 1201700324, 1208700044, 1213100124, 1201700334, 1208700054, 1213100134, 1201700344, 1209400164, 1213100144, 1201700354, 1209400174, 1213100154, 1201700364, 1209400184,1206500014, 1209400194, 1206500024, 1209400204, 1206500034, and 1210900264.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1880-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1880-2013. Report a data issue: hello@regdatalab.com.