Z-1880-2017 Class II Terminated

Recalled by Savaria Concord Lifts, Inc. — Brampton, N/A

Recall Details

Product Type
Devices
Report Date
May 3, 2017
Initiation Date
April 12, 2017
Termination Date
October 16, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
74

Product Description

Omega Incline Platform Lift

Reason for Recall

Savaria Concord Lifts Inc has decided to initiate this recall of its Omega Incline Platform Lift because bolts can come loose causing the carriage to separate from the trolley and the unit to fail

Distribution Pattern

Worldwide Distribution - US Distribution and to North America.

Code Information

s/n or lot number - Job Number 630046 through 630119