Z-1880-2018 Class II Terminated
Recalled by Visaris DOO
— Belgrade
FDA device recall Z-1880-2018 was initiated by Visaris DOO on Apr 5, 2018 and is designated Class II. Reason for recall: A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant… The recall status is terminated (terminated Sep 19, 2019). Affected quantity: 8 devices (US), 14 devices (international).
Recall details
- Product Type
- Devices
- Report Date
- May 23, 2018
- Initiation Date
- Apr 5, 2018
- Termination Date
- Sep 19, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8 devices (US), 14 devices (international)
Product description
Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.2 The Vision systems are intended for use in all areas of diagnostic radiography imaging and specifically at secondary and tertiary levels of medical care. Powerful positioning capability makes this system particularly suited for institutions with immobile or hardly mobile patients, such as trauma wards and emergency departments. The system s versatility also makes it suitable for unified diagnostic imaging services for medium size hospitals where this system can cover all diagnostic radiography needs. In general, Vision systems are intended to position the patient, x-ray source and digital detector in such a way as to expose only the desired patient anatomy to X-rays and capture the resulting x-ray image. They are then intended to process the acquired image to a diagnostic viewing level and export it in DICOM format through an option of several distribution media including film, paper, CD-DVD and digital PACS archives. The system is intended for routine, daily, dedicated use in radiographic diagnosis. It can be used in any type of specialized radiology wards or for general radiographic diagnostics
Reason for recall
A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.
Distribution pattern
Worldwide Distribution - US Distribution to the states of : FL, LA, MD, MS, NJ, and NY., and to the countries of : Czech Republic, Germany, Serbia, Switzerland, Taiwan, and United Kingdom.
Code information
System SN VC16B-02002, Column SN BA-215-1800-862-1-30008; System SN VC15L-02002, Column SN BA-215-1800-862-1-30015; System SN VC15L-01002, Column SN BA-215-1800-862-1-30011; System SN VC15J-01002, Column SN BA-215-1500-792-10001; System SN VC15I-02001, Column SN BA-215-1800-862-1-30006
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1880-2018
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Visaris DOO? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Visaris DOO: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Visaris DOO →
Use it in ChatGPT
Get a free API key
How the connector works