Z-1880-2018 Class II Terminated

Recalled by Visaris DOO — Belgrade

FDA device recall Z-1880-2018 was initiated by Visaris DOO on Apr 5, 2018 and is designated Class II. Reason for recall: A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant… The recall status is terminated (terminated Sep 19, 2019). Affected quantity: 8 devices (US), 14 devices (international).

Recall details

Product Type
Devices
Report Date
May 23, 2018
Initiation Date
Apr 5, 2018
Termination Date
Sep 19, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 devices (US), 14 devices (international)

Product description

Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.2 The Vision systems are intended for use in all areas of diagnostic radiography imaging and specifically at secondary and tertiary levels of medical care. Powerful positioning capability makes this system particularly suited for institutions with immobile or hardly mobile patients, such as trauma wards and emergency departments. The system s versatility also makes it suitable for unified diagnostic imaging services for medium size hospitals where this system can cover all diagnostic radiography needs. In general, Vision systems are intended to position the patient, x-ray source and digital detector in such a way as to expose only the desired patient anatomy to X-rays and capture the resulting x-ray image. They are then intended to process the acquired image to a diagnostic viewing level and export it in DICOM format through an option of several distribution media including film, paper, CD-DVD and digital PACS archives. The system is intended for routine, daily, dedicated use in radiographic diagnosis. It can be used in any type of specialized radiology wards or for general radiographic diagnostics

Reason for recall

A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.

Distribution pattern

Worldwide Distribution - US Distribution to the states of : FL, LA, MD, MS, NJ, and NY., and to the countries of : Czech Republic, Germany, Serbia, Switzerland, Taiwan, and United Kingdom.

Code information

System SN VC16B-02002, Column SN BA-215-1800-862-1-30008; System SN VC15L-02002, Column SN BA-215-1800-862-1-30015; System SN VC15L-01002, Column SN BA-215-1800-862-1-30011; System SN VC15J-01002, Column SN BA-215-1500-792-10001; System SN VC15I-02001, Column SN BA-215-1800-862-1-30006

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1880-2018
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.