Z-1880-2024 Class I Ongoing

Recalled by ARROW INTERNATIONAL Inc. — Morrisville, NC

FDA device recall Z-1880-2024 was initiated by ARROW INTERNATIONAL Inc. on Apr 29, 2024 and is designated Class I. Reason for recall: Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the… The recall status is ongoing. Affected quantity: 3138 units.

Recall Details

Product Type
Devices
Report Date
Jun 12, 2024
Initiation Date
Apr 29, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3138 units

Product Description

Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U

Reason for Recall

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

Distribution Pattern

Worldwide distribution. US nationwide including Puerto Rico, AU, BA, BE, BG, BN, CA, CL, CN, CO, CY, CZ, DE, EC, ES, FR, GB, GP, GR, HU, HK, IE, IL, IN, IQ, IT, JO, JP, KW, KZ, LB, LU, MA, MN, MK, MX, MY, NL, OM, PA, PF, PH, PL, PT, QA, RE, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UZ, ZA, and UAE.

Code Information

a) REF IAB-05830-U, UDI: (01)10801902172850(17)240531(10)18F22F0023, (01)10801902172850(17)240630(10)18F22G0005, (01)10801902172850(17)240630(10)18F22G0010, (01)10801902172850(17)240731(10)18F22H0033, (01)10801902172850(17)250831(10)18F23J0031, (01)10801902172850(17)250930(10)18F23K0036, (01)10801902172850(17)250930(10)18F23K0042, (01)10801902172850(17)251130(10)18F23M0001, (01)10801902182019(17)250430(10)18F23E0032, (01)10801902182019(17)250831(10)18F23J0002, (01)10801902182019(17)250831(10)18F23J0082, (01)10801902182019(17)250930(10)18F23K0025, (01)10801902182019(17)260331(10)18F24C0131; b) REF IAB-05840-U, UDI: (01)10801902161922(17)240630(10)18F22G0051, (01)10801902161922(17)241231(10)18F23A0002, (01)10801902161922(17)241231(10)18F23A0021, (01)10801902161922(17)241231(10)18F23A0060, (01)10801902161922(17)250731(10)18F23G0068, (01)10801902161922(17)250930(10)18F23K0012, (01)10801902161922(17)251031(10)18F23K0053, (01)10801902161922(17)251031(10)18F23K0063, (01)10801902161922(17)260131(10)18F24B0051, (01)10801902172874(17)240430(10)18F22E0012, (01)10801902172874(17)240630(10)18F22G0011, (01)10801902172874(17)240630(10)18F22G0016, (01)10801902172874(17)250930(10)18F23K0039, (01)10801902172874(17)251031(10)18F23L0010, (01)10801902172874(17)251031(10)18F23L0031, (01)10801902172874(17)251031(10)18F23L0039, (01)10801902172874(17)251130(10)18F23K0008

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1880-2024
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.