Z-1880-2024 Class I Ongoing
Recalled by ARROW INTERNATIONAL Inc.
— Morrisville, NC
FDA device recall Z-1880-2024 was initiated by ARROW INTERNATIONAL Inc. on Apr 29, 2024 and is designated Class I. Reason for recall: Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the… The recall status is ongoing. Affected quantity: 3138 units.
Recall details
- Product Type
- Devices
- Report Date
- Jun 12, 2024
- Initiation Date
- Apr 29, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3138 units
Product description
Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U
Reason for recall
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.
Distribution pattern
Worldwide distribution. US nationwide including Puerto Rico, AU, BA, BE, BG, BN, CA, CL, CN, CO, CY, CZ, DE, EC, ES, FR, GB, GP, GR, HU, HK, IE, IL, IN, IQ, IT, JO, JP, KW, KZ, LB, LU, MA, MN, MK, MX, MY, NL, OM, PA, PF, PH, PL, PT, QA, RE, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UZ, ZA, and UAE.
Code information
a) REF IAB-05830-U, UDI: (01)10801902172850(17)240531(10)18F22F0023, (01)10801902172850(17)240630(10)18F22G0005, (01)10801902172850(17)240630(10)18F22G0010, (01)10801902172850(17)240731(10)18F22H0033, (01)10801902172850(17)250831(10)18F23J0031, (01)10801902172850(17)250930(10)18F23K0036, (01)10801902172850(17)250930(10)18F23K0042, (01)10801902172850(17)251130(10)18F23M0001, (01)10801902182019(17)250430(10)18F23E0032, (01)10801902182019(17)250831(10)18F23J0002, (01)10801902182019(17)250831(10)18F23J0082, (01)10801902182019(17)250930(10)18F23K0025, (01)10801902182019(17)260331(10)18F24C0131; b) REF IAB-05840-U, UDI: (01)10801902161922(17)240630(10)18F22G0051, (01)10801902161922(17)241231(10)18F23A0002, (01)10801902161922(17)241231(10)18F23A0021, (01)10801902161922(17)241231(10)18F23A0060, (01)10801902161922(17)250731(10)18F23G0068, (01)10801902161922(17)250930(10)18F23K0012, (01)10801902161922(17)251031(10)18F23K0053, (01)10801902161922(17)251031(10)18F23K0063, (01)10801902161922(17)260131(10)18F24B0051, (01)10801902172874(17)240430(10)18F22E0012, (01)10801902172874(17)240630(10)18F22G0011, (01)10801902172874(17)240630(10)18F22G0016, (01)10801902172874(17)250930(10)18F23K0039, (01)10801902172874(17)251031(10)18F23L0010, (01)10801902172874(17)251031(10)18F23L0031, (01)10801902172874(17)251031(10)18F23L0039, (01)10801902172874(17)251130(10)18F23K0008
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1880-2024
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for ARROW INTERNATIONAL Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for ARROW INTERNATIONAL Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about ARROW INTERNATIONAL Inc. →
Use it in ChatGPT
Get a free API key
How the connector works