Z-1881-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
April 29, 2026
Initiation Date
February 27, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
270,311 total

Product Description

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes BARIATRIC DYNJ905153M, DYNJ905153N GENERAL LAPAROSCOPY PACK DYNJ81602 LAPAROSCOPY DYNJ905157K, DYNJ905157L LAPAROTOMY PACK DYNJ46675G MAJOR LITHOTOMY PACK DYNJ81607 MAJOR PLUS PACK DYNJ905160O MINOR GENERAL PACK DYNJ81599

Reason for Recall

Unapproved design changes to the products outside of the 510(k) clearance.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.

Code Information

DYNJ905153M UDI-DI 10195327556761 (EA) 40195327556762 (CS) LOT 24CMG485 DYNJ905153N UDI-DI 10198459017797 (EA) 40198459017798 (CS) LOT 25DMC254 DYNJ81602 UDI-DI 10195327123857 (EA) 40195327123858 (CS) LOT 24ABP080 DYNJ905157K UDI-DI 10195327556921 (EA) 40195327556922 (CS) LOTS 24DMF349 24EMC297 DYNJ905157L UDI-DI 10198459017711 (EA) 40198459017712 (CS) LOTS 24FMD083 24HMD400 24IME332 24JMA234 24KMG588 25AMA393 25AMJ541 25BMI202 25CMD909 25CMH371 DYNJ46675G UDI-DI 10195327419066 (EA) 40195327419067 (CS) LOT 24ABS814 DYNJ81607 UDI-DI 10195327123918 (EA) 40195327123919 (CS) LOT 24ABP097 DYNJ905160O UDI-DI 10198459017742 (EA) 40198459017743 (CS) LOTS 24HMG093 24IME714 24JMI924 24LMB556 25AME613 25BMJ076 DYNJ81599 UDI-DI 10195327124021 (EA) 40195327124022 (CS) LOT 24BBB847 ***Updated 5/20/26 - The following lots were included in the customer letter but not in the FDA submission.*** DYNJ905153M lots 24BMH968 24EMG682; DYNJ905153N lots 24HMG087 24IMA442 24IMG245; DYNJ905157K lots 23LME674 24AMH261 24AMJ205 24BME886 24CMD057