Z-1883-2013 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

FDA device recall Z-1883-2013 was initiated by Siemens Healthcare Diagnostics, Inc on May 10, 2013 and is designated Class II. Reason for recall: A rare variant of TSH, identified in a small cluster of patients, is not detected. The recall status is terminated (terminated May 10, 2016). Affected quantity: 33506.

Recall details

Product Type
Devices
Report Date
Aug 14, 2013
Initiation Date
May 10, 2013
Termination Date
May 10, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33506

Product description

Dimension EXL LOCI Module TSHL Flex Reagent Cartridge (REF RF612). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin).

Reason for recall

A rare variant of TSH, identified in a small cluster of patients, is not detected.

Distribution pattern

Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.

Code information

Per recall strategy.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1883-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Healthcare Diagnostics, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Healthcare Diagnostics, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Healthcare Diagnostics, Inc →

Source: openFDA record for Z-1883-2013. Report a data issue: hello@regdatalab.com.