Z-1888-2020 Class II Ongoing

Recalled by Inpeco S.A. — Lugano, N/A

FDA device recall Z-1888-2020 was initiated by Inpeco S.A. on Apr 7, 2020 and is designated Class II. Reason for recall: The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge. The recall status is ongoing. Affected quantity: 23 units.

Recall details

Product Type
Devices
Report Date
May 13, 2020
Initiation Date
Apr 7, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23 units

Product description

Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf

Reason for recall

The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.

Distribution pattern

Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri

Code information

Serial Numbers: ACP.357, ACP.388, ACP.396, ACP.397, ACP.399, ACP.400, ACP.402, ACP.404, ACP.405, ACP.407, ACP.408, ACP.409, ACP.410, ACP.412, ACP.413, ACP.416, ACP.417, ACP.418, ACP.419, ACP.420, ACP.421, ACP.422, ACP.423

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1888-2020
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Inpeco S.A.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1888-2020. Report a data issue: hello@regdatalab.com.