Z-1891-2020 Class II Ongoing

Recalled by Intuitive Surgical, Inc. — Sunnyvale, CA

Recall Details

Product Type
Devices
Report Date
May 13, 2020
Initiation Date
July 1, 2019
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

daVinci XI Sureform 60 Reload, 60 mm black 6-Row 4.6 mm, Ref: 48360T-08 , UDI: 00886874113851, Rx only, SterileEO - Product Usage: are intended to be used with da Vinci Surgical Systems for resection, transection, and, or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.

Reason for Recall

The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.

Distribution Pattern

Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Switzerland, United Kingdom.

Code Information

Lot numbers: T10180615; T10180924; T10181003; T11181003; T10181010; T10181016; T10181029; T10181107; T10181109; T10181113