Z-1891-2020 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 13, 2020
- Initiation Date
- July 1, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
daVinci XI Sureform 60 Reload, 60 mm black 6-Row 4.6 mm, Ref: 48360T-08 , UDI: 00886874113851, Rx only, SterileEO - Product Usage: are intended to be used with da Vinci Surgical Systems for resection, transection, and, or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.
Reason for Recall
The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line. Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.
Distribution Pattern
Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Switzerland, United Kingdom.
Code Information
Lot numbers: T10180615; T10180924; T10181003; T11181003; T10181010; T10181016; T10181029; T10181107; T10181109; T10181113