Z-1897-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1897-2019 was initiated by Siemens Medical Solutions USA, Inc on June 6, 2019 and is designated Class II. Reason for recall: Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to th… The recall status is terminated (terminated February 9, 2021). Affected quantity: 29.

Recall Details

Product Type
Devices
Report Date
July 10, 2019
Initiation Date
June 6, 2019
Termination Date
February 9, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
29

Product Description

Siemens Artis zeego--Interventional Fluoroscopic X-Ray System Model Number: 10280959

Reason for Recall

Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Distribution Pattern

Nationwide

Code Information

Serial Numbers: 160335 160430 160368 160469 160357 160370 160369 160509 160427 160433 160443 160428 160429 160463 160001 160461 160486 160800 160487 160488 160507 160840 160862 160856 160823 160914 160930 161235 160401