Z-1897-2020 Class II Terminated
FDA device recall Z-1897-2020 was initiated by Intuitive Surgical Inc on Apr 11, 2019 and is designated Class II. Reason for recall: Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking. The recall status is terminated (terminated Apr 8, 2024). Affected quantity: 270 USMs on a combination of 100 unique systems (4 USM per system).
Recall details
- Product Type
- Devices
- Report Date
- May 13, 2020
- Initiation Date
- Apr 11, 2019
- Termination Date
- Apr 8, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 270 USMs on a combination of 100 unique systems (4 USM per system)
Product description
Reason for recall
Distribution pattern
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain, Sweden, and the United Kingdom.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1897-2020
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
Need the full FDA record for Intuitive Surgical Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.