Z-1897-2020 Class II Terminated
FDA device recall Z-1897-2020 was initiated by Intuitive Surgical Inc on April 11, 2019 and is designated Class II. Reason for recall: Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking. The recall status is terminated (terminated April 8, 2024). Affected quantity: 270 USMs on a combination of 100 unique systems (4 USM per system).
Recall Details
- Product Type
- Devices
- Report Date
- May 13, 2020
- Initiation Date
- April 11, 2019
- Termination Date
- April 8, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 270 USMs on a combination of 100 unique systems (4 USM per system)
Product Description
Reason for Recall
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain, Sweden, and the United Kingdom.