Z-1897-2020 Class II Terminated

Recalled by Intuitive Surgical Inc — Sunnyvale, CA

FDA device recall Z-1897-2020 was initiated by Intuitive Surgical Inc on April 11, 2019 and is designated Class II. Reason for recall: Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking. The recall status is terminated (terminated April 8, 2024). Affected quantity: 270 USMs on a combination of 100 unique systems (4 USM per system).

Recall Details

Product Type
Devices
Report Date
May 13, 2020
Initiation Date
April 11, 2019
Termination Date
April 8, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
270 USMs on a combination of 100 unique systems (4 USM per system)

Product Description

da Vinci X Surgical System, IS4200 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the IS4000 system and its primary function is to support the instrument arms and camera arm. The system has four Universal Surgical Manipulators (USM) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. There are four USM per da Vinci X surgical system.

Reason for Recall

Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain, Sweden, and the United Kingdom.

Code Information

UDI: 00886874114216 Device Listing: D295931 Serial Numbers: SL0085, SL0315, SL0317, SL0324, SL0338, SL0350, SL0351, SL0353, SL0354, SL0357, SL0369