Z-1898-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 10, 2019
- Initiation Date
- June 6, 2019
- Termination Date
- February 9, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 12
Product Description
Siemens Artis Q floor --Interventional Fluoroscopic X-Ray System Model Number: 10848280
Reason for Recall
Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel
Distribution Pattern
Nationwide
Code Information
Serial Numbers: 103008 103030 103028 103258 103221 103222 103277 103220 103279 103285 103286 103295