Z-1898-2025 Class II Ongoing

Recalled by Silex Medical, LLC — Southington, CT

FDA device recall Z-1898-2025 was initiated by Silex Medical, LLC on Jan 9, 2024 and is designated Class II. Reason for recall: Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel. The recall status is ongoing. Affected quantity: 7 units.

Recall details

Product Type
Devices
Report Date
Jun 11, 2025
Initiation Date
Jan 9, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product description

Signature Laparoscopic Instruments, Alligator Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5863050.

Reason for recall

Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

Distribution pattern

US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.

Code information

Model/REF Number: 5863050. UDI-DI: B33158630500. Lot Number: M28186. Serial Numbers: 1051, 1052, 1053, 1054, 1055, 1056, 1057, 1058, 1059

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1898-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Silex Medical, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Silex Medical, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Silex Medical, LLC →

Source: openFDA record for Z-1898-2025. Report a data issue: hello@regdatalab.com.