Z-1899-2020 Class II Terminated

Recalled by Varian Medical Systems, Inc. — Palo Alto, CA

Recall Details

Product Type
Devices
Report Date
May 13, 2020
Initiation Date
August 12, 2019
Termination Date
January 21, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
151 units

Product Description

Varian Multileaf Collimator [MLC]; Millenium Multileaf Collimator [52, 80, 120 leaf]; High Definition 120 Multileaf Collimator Model Number(s): product code: H54 and H56; MLC software version 8.5 - Product Usage: is intended to assist the clinician in the delivery of external beam radiation to defined target volumes during radiosurgery and radiotherapy while sparing surrounding normal tissue and critical organs from excess radiation.

Reason for Recall

After a recent upgrade to the collimator software version 8.5, the firm became aware that the multi-leaf collimator leaves did not move during an arc treatment. The issue is related to a transient carriage primary- secondary interlock due to carriage fault on the MLC that immediately preceded the initiation of the arc treatment. The issue will occur only with MLC software version 8.5 and only affects conformal arc treatments, including VMAT and RapidArc. There have been no reports of adverse health consequences due to this issue.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of TN, FL, IN, PA, NV, WA, DC, CA, AL, IL, MD, VA, TX, MN, AR, GA, OR, MS, MO, NY, SC, DE, AZ, NH, MI, NC, OH, PA and OK. The country of Canada.

Code Information

Model Number(s): product code: H54 and H56; MLC software version 8.5. UDIs: HD 120 MLC Head Assembly. UDI: 00899475002318 80 Leaf Assy, MLC Head, UDI: 00899475002325 120 Leaf Assy, MLC Head, UDI: 00899475002332