Z-0771-2019 Class II Terminated
Recalled by Varian Medical Systems, Inc.
— Palo Alto, CA
FDA device recall Z-0771-2019 was initiated by Varian Medical Systems, Inc. on Dec 26, 2018 and is designated Class II. Reason for recall: There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range… The recall status is terminated (terminated Mar 10, 2021). Affected quantity: 1450 devices.
Recall details
- Product Type
- Devices
- Report Date
- Jan 30, 2019
- Initiation Date
- Dec 26, 2018
- Termination Date
- Mar 10, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1450 devices
Product description
Eclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, elec- tron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Reason for recall
There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) X<Y: R <R (R is too small) and 3) X>Y R >R (R is too big).
Distribution pattern
Worldwide distribution: US (nationwide) and countries of: Australia, Brazil, Denmark France, Germany, Japan, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Sweden, Switzerland, Taiwan, and United Kingdom.
Code information
Versions 13.X and 15.X
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0771-2019
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Varian Medical Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works