Z-2150-2016 Class II Terminated

Recalled by Varian Medical Systems, Inc. — Palo Alto, CA

FDA device recall Z-2150-2016 was initiated by Varian Medical Systems, Inc. on Jun 23, 2016 and is designated Class II. Reason for recall: Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user… The recall status is terminated (terminated Aug 24, 2018). Affected quantity: 21 AlignRT Plus in US, 1 International.

Recall details

Product Type
Devices
Report Date
Jul 20, 2016
Initiation Date
Jun 23, 2016
Termination Date
Aug 24, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21 AlignRT Plus in US, 1 International.

Product description

Optical Surface Monitoring System{Align RT Plus} This is used on Varian's TrueBeam EDGE; Radiology: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to Withhold the beam automatically during radiation delivery.

Reason for recall

Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different couch rotation from the plan couch rotation while performing patient alignment with Optical Surface Monitoring System [OSMS].

Distribution pattern

Worldwide Distribution: US (nationwide) and countries of: Austria, Azerbaijan, Brunei Darussalam, Bulgaria, Denmark, Georgia, Germany, India, Italy, Malaysia, Morocco, Norway, Portugal, Switzerland, and United Kingdom.

Code information

TrueBeam version 2.0/2.5 with any version OSMS. AlignRT Serial Numbers: 247-0200 247-0209 248-0013 248-0057 248-0070 249-0057 249-0066 249-0163 249-0164 249-0195 249-0226 249-0274 249-0050 249-0286 249-0277 248-0044 249-0227 247-0208 249-0047 247-0218 248-0050 248-0043 248-0048 249-0293 249-0155.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2150-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Varian Medical Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2150-2016. Report a data issue: hello@regdatalab.com.